Comparison of rebound tonometry with applanation tonometry for intraocular pressure measurement

Journal Title: Indian Journal of Clinical and Experimental Ophthalmology - Year 2017, Vol 3, Issue 3

Abstract

Purpose: To compare the readings of the Goldmann’s applanation tonometry with that of the iCare Ta01i in Indian patients. Materials and Methods: Intraocular pressure measurements were obtained in 440 eyes of 220 subjects in which 320 were normal eyes and 120 eyes had a high IOP. A thorough examination was done and IOP was measured in all subjects using iCare Ta01 tonometer and the Goldmann’s applanation tonometer 3 times each and the average was taken. All examinations and measurements were done by the same 2 personnel. Results: Intraocular pressure measurements with ICare Ta01 tonometer compared with the Goldmann’s applanation tonometer were not found to have a significant difference (mean ±2.34mmHg). The 95% confidence interval between readings with the 2 methods was within acceptable limits. Conclusion: The ICare Ta01 tonometer measurement of intraocular pressure showed excellent correlation with those from the Goldmann applanation tonometer, which is the gold standard for IOP measurement. The iCare, with all its advantages over the GAT, can be used as a routine instrument for measurement of IOP in ophthalmology practice.

Authors and Affiliations

CN Madhusudhana, S Shyam

Keywords

Related Articles

Effect of Type II Diabetes mellitus on Intra ocular pressure in Central India

Aim: Intraocular pressure distribution and its association with Type 2 Diabetes Mellitus. Material and Methods: A comparative cross-sectional study was done in a tertiary care hospital. Patients of 20-80 years of age whe...

Visual outcome of hydrophilic and hydrophobic foldable posterior chamber intraocular lenses in cataract surgery

Introduction: As there is no pharmacological treatment for cataract, the standard treatment is surgical removal of the opacified lens and implantation of an artificial intraocular lens (IOL). The standard method of catar...

Correlation of diabetic retinopathy with hypertension

Diabetes mellitus is internationally on a pandemic. India is the diabetic capital of the world. With growing stress and life style changes hypertension too is on the rise. The same group of patients are afflicted by both...

The intraocular pressure changes following intravitreal bevacizumab injections in an Indian population

Aimnbsp To assess the changes in Intraocular pressure IOP after intravitreal bevacizumab therapyDesign Prospective Observational Tertiary care hospital based studyParticipants 95 patients receiving intravitreal bevacizum...

Comparison of postoperative cystoid macular edema in manual SICS vs phacoemulsification

Aim: To compare the macular edema occurring in patients postoperatively in uncomplicated manual SICS and uncomplicated phacoemulsification on the basis of macular thickness using optical coherence tomography (OCT). Setti...

Download PDF file
  • EP ID EP300641
  • DOI 10.18231/2395-1451.2017.0081
  • Views 73
  • Downloads 0

How To Cite

CN Madhusudhana, S Shyam (2017). Comparison of rebound tonometry with applanation tonometry for intraocular pressure measurement. Indian Journal of Clinical and Experimental Ophthalmology, 3(3), 329-332. https://www.europub.co.uk/articles/-A-300641