THE DEPENDENCE OF BRASS HARDNESS MEASUREMENT SYSTEM ON TIME

Journal Title: Metrology and Measurement Systems - Year 2007, Vol 14, Issue 4

Abstract

Measurement system analysis (MSA) was used for estimation of hardness (HBS) measure- ment system capability. The time dependence of capability was evaluated. Experimental material Cu-Zn-Al cast alloy was used as shape memory alloy. The average and range method with 2 appraisers, 10 samples and 3 trials (indentations). was used to quantify repeatability (% EV), reproducibility (% AV) and part variation (% PV - variation between samples). The measurement was repeated 5 times in intervals of 7 days. As the values of %GRR are between 89.4% and 63.7% and values of "ndc" between 0.71 and 1.71 the used hardness measurement system is not capable and it depends heavily on the long-term variability.

Authors and Affiliations

JOZEF PETRÍK, VOJTECH ŠPETUCH

Keywords

Related Articles

COMPARISON BETWEEN SIMPLE POLYNOMIALS AND AN ITS-90 BASED REFERENCE FUNCTIONS OF EMF VS. TEMPERATURE RELATION FOR HIGH PURITY NOBLE METAL THERMOCOUPLES

Emf versus temperature reference polynomial functions covering the temperature range from 100° C up to 960° C have been developed for three types of noble metal thermocouples, which are Pt-10%Rh/Pt, Au/Pt and Pt/Pd. Our...

INFLUENCE OF SELECTED FACTORS FOR LOCALIZATION ACCURACY IN THE OPTICAL VORTEX INTERFEROMETER

Optical vortices were intensively studied during the last decade. In the literature there are papers presenting applications of the optical vortices. The regular net of optical vortices generated by the three plane waves...

SEVERAL APPROACHES TO ADC TRANSFER FUNCTION APPROXIMATION AND THEIR APPLICATION FOR ADC NON-LINEARITY CORRECTION

The performance of current electronic devices is mostly limited by analog front-end and analog-to-digital converter's (ADC) actual parameters. One of the most important parameters is ADC nonlinearity. The correction of t...

THE EFFECT OF DOUBLE VENTURI'S STRUCTURAL PARAMETERS ON THE REINFORCEMENT VALUE AND FLOW COEFFICIENT

Reinforcement factor of a double Venturi is one of the more important parameters related to sensor shape and structure. The evaluation and the effect of the double Venturi's structural parameters on its value were carrie...

NEW CONCEPT OF A MEASUREMENT PROBE FOR HIGH IMPEDANCE SPECTROSCOPY

The paper presents a new measurement probe working as the input circuit for a high impedance spectroscopy analyzer. The probe allows to measure impedance in the range of 100 Ω < |Zx | < 100 GΩ. The probe extracts t...

Download PDF file
  • EP ID EP83141
  • DOI -
  • Views 108
  • Downloads 0

How To Cite

JOZEF PETRÍK, VOJTECH ŠPETUCH (2007). THE DEPENDENCE OF BRASS HARDNESS MEASUREMENT SYSTEM ON TIME. Metrology and Measurement Systems, 14(4), 595-602. https://www.europub.co.uk/articles/-A-83141