Experimental Assessment of a Laboratory Rotary Sample Splitter

Abstract

A binary mixture of iron chips and sand was used to experimentally test the performance of a constructed laboratory sample splitter. Four tests were conducted and six subsamples were produced in each case. The subsamples were separated into their constituents and the percentage weight of iron chips was determined. The results obtained were examined statistically and it was observed that the mean value of the percentage weight of iron chips in the subsamples range from 30.4 – 35.7% which compared well with the sample true value of 33.3%. The biases of the subsamples range from 0.2 – 2.9%, the variance ranges from 0.790 – 2.334 while the coefficient of variation is 0.025 – 4.6%. The standard deviation of the subsamples produced in the experiment is 0.0889 – 1.5277%. The level of accuracy and precision of rotary sample riffling is superior to some other sampling techniques and the machine tested gave an intermediate performance indicating that it is suitable for routine and non-critical works. Rotary Sample Splitter is therefore recommended for splitting dry granular samples, whenever an efficient, accurate and mechanized technique for sample splitting is desired.

Authors and Affiliations

Nenuwa , Olushola B

Keywords

Related Articles

Design Analysis of Slotted Wideband Rectangular Micro-Strip Patch Antenna

In this paper a small slotted micro-strip patch antenna. The design and simulation has done for efficiently perform between4.5GHz to 5.8GHz. In a conventional rectangular patch, a 45 degree cut has introduced in three di...

Structural and Ionic Conductivity Properties of Gamma Irradiated HPMC Polymer Electrolyte Films

A solid biopolymer electrolytes (BEs) based on hydroxypropyl methylcellulose (HPMC) doped with different concentrations of cadmium chloride (CdCl2 ) and ceric ammonium nitrate (NH4 ) 2 Ce(NO3 ) 6 (CAN) were prepared by s...

A Review on Fabrication Methods of Porous Titanium Dental Implant

Titanium has got excellent mechanical and biological properties. Itis nearly inert biologically as due to its biocompatibility. But titanium and its alloys have a high Young’s modulus compared to the bone. To improve oss...

TiO2 Nanotube Arrays formation in Titanium Dental Implants by Anodization

Dental problems are increasing day by day. Even today, we don’t have a permanent solution for the replacement of missing tooth. Dentures, bridges and dental implants are being used to solve the above problem. Due to vari...

‘Cast in India’-A Sequel of “Make in India”

Make in India, a core policy initiative, devised as on Sept. 2014 was initiated as a potent prescription to recharge the ailing manufacturing sector, which was further threatened with increasing capital imports and also...

Download PDF file
  • EP ID EP374515
  • DOI -
  • Views 125
  • Downloads 0

How To Cite

Nenuwa, Olushola B (2018). Experimental Assessment of a Laboratory Rotary Sample Splitter. Journal of Advanced Research in Manufacturing, Material Science & Metallurgical Engineering, 5(1), 15-21. https://www.europub.co.uk/articles/-A-374515